Walk In Stability Chamber, Stability Chambers, Humidity Chamber, Bacteriological Incubator, Mumbai, India
Walk In Stability Chamber

Walk - In Stability Chamber

Temp. range : 10șC above ambient to 60șC,10șC to 60șC
Accuracy : ± 0.1șC to 0.5șC
Humidity range : 25% RH to 98% RH Accuracy : ± 3% RH.

Application:
These chambers are specially designed to carry out stability & shelf life TESTS on drugs & drugs substance according to ICH Guidelines & are mainly used in shelf life studies, conditioning of samples, stability tests for drugs & pharmaceutical Industries

Construction :
Pre-fabricated modular panel with PUF insulation enabling to assemble it on site, light weight & easy to handle. All the edges are molded tight fit which prevents condensation & temperature losses, stainless steel trays & trollies.

Air Circulation :
A high efficient fan motor with blower ensures uniform temp.

Heating & Humidification :
Air heaters are provided for heating & maintaining above ambient temperatures. A water boiler with reservoir & float valve produces steam for Humidity.


Refrigeration System :
A hermetically sealed compressor with CFC free refrigerant & optional stand by refrigeration System & Humidity System for uninterrupted operation. The both the systems can be operated alternately for a certain period, thus increasing the life
of the system & in case of break down the alternate stand by system can be operated for uninterrupted services

Temp. Control :
Microprocessor based PID Temp. Control are used for accurate control of temp. & Humidity. Printer interface facility with data storage to print records with variable print time intervals. RS-485 interface to personal computer for multiple
networking of chambers, & also confirms to U. S FDA guidelines. Chamber illumination with fluorescent or UV lights. Cyclic timer for regulating illumination conditions 16 point (8+8) temp. & Humidity Scanner for mapping purpose.

 

STI – A

230 x 110 x 110

2800

STI – B

200 x 165 x 200

6500

STI – C

200 x 215 x 200

8500

STI – D

200 x 265 x 200

10500

STI – E

250 x 300 x 200

15000

STI – F

250 x 400 x 200

20000

Due to continuous development & improvements in design, we reserve the right to change the specification without notice.

Validation Protocols :
Validation Protocols as per DQ, IQ, OQ & PQ, chamber validation with & without load, with complete documentation & certification traceable to national & International standards.

Optional :
Door Access systems with swipe card & software to authorize & monitor door accessibility. Safety circuit to cut off the whole system in case of any malfunction in the chamber (Only applicable in microprocessor based models.)

Built in temp. Deviation audio/ visual alarms. safety thermostat prevent overshooting  of temp. HRC fuses for compressor, heaters & mains

Special Care :
Suitable KVA Servo Controlled Voltage Stabilizers is recommended
 