Walk -
In Stability Chamber
Temp.
range : 10șC above ambient to 60șC,10șC to 60șC
Accuracy : ± 0.1șC to 0.5șC
Humidity range : 25% RH to 98% RH Accuracy : ± 3% RH.
Application:
These chambers are specially designed to carry out
stability & shelf life TESTS on drugs & drugs
substance according to ICH Guidelines & are mainly
used in shelf life studies, conditioning of samples,
stability tests for drugs & pharmaceutical
IndustriesConstruction
:
Pre-fabricated modular panel with PUF insulation
enabling to assemble it on site, light weight & easy to
handle. All the edges are molded tight fit which
prevents condensation & temperature losses, stainless
steel trays & trollies.
Air Circulation :
A high efficient fan motor with blower ensures uniform
temp. |
Heating & Humidification :
Air heaters are provided for heating & maintaining above
ambient temperatures. A water boiler with reservoir &
float valve produces steam for Humidity.
Refrigeration System :
A hermetically sealed compressor with CFC free
refrigerant & optional stand by refrigeration System &
Humidity System for uninterrupted operation. The both
the systems can be operated alternately for a certain
period, thus increasing the life
of the system & in case of break down the alternate
stand by system can be operated for uninterrupted
services
Temp. Control :
Microprocessor based PID Temp. Control are used for
accurate control of temp. & Humidity. Printer interface
facility with data storage to print records with
variable print time intervals. RS-485 interface to
personal computer for multiple
networking of chambers, & also confirms to U. S FDA
guidelines. Chamber illumination with fluorescent or UV
lights. Cyclic timer for regulating illumination
conditions 16 point (8+8) temp. & Humidity Scanner for
mapping purpose.
|
|
TECHNICAL
SPECIFICATIONS.
|
Model In |
Internal Dimensions (cms) |
Capacity In Ltrs |
|
STI – A |
230 x 110 x 110 |
2800 |
|
STI – B |
200 x 165 x 200 |
6500 |
|
STI – C |
200 x 215 x 200 |
8500 |
|
STI – D |
200 x 265 x 200 |
10500 |
|
STI – E |
250 x 300 x 200 |
15000 |
|
STI – F |
250 x 400 x 200 |
20000 |
Due to continuous development &
improvements in design, we reserve the right to change
the specification without notice.
|
Validation Protocols :
Validation Protocols as per DQ, IQ, OQ & PQ, chamber
validation with & without load, with complete
documentation & certification traceable to national &
International standards.
Optional
:
Door Access systems with swipe card & software to
authorize & monitor door accessibility. Safety circuit
to cut off the whole system in case of any malfunction
in the chamber (Only applicable in microprocessor based
models.)
Safety Features
:
Built in temp.
Deviation audio/ visual alarms. safety thermostat prevent overshooting
of temp. HRC fuses for compressor, heaters & mains
Special Care :
Suitable KVA Servo Controlled Voltage Stabilizers is recommended |